O3 Analyzer

O3 Analyzer美智

Technical principle

UV adsorption Ozone analyzer is developed based on Beer-Lambert technology. It is applied to monitor low concentration O3 in ambient air. I = Io e-αLc 254nm UV light passes through sample gas cell and is absorbed by ozone inside it. Adsorption value has a certain ration with O3 concentration. By cyclically alternating of magnetic valve and measuring sample gas and the sample without O3, the true and reliable measured concentration of ozone is obtained at last.

Measurement range

0~500ppb, 0~10ppm(optional)

Unit

nmol/mol (ppb), μmol/mol (ppm), μg/m3, mg/m3(optional)

Zero noise

<0.5 ppb(RMS)

Detection lower limit

1 ppb

Zero drift 

≤1 ppb / 24 hours

Span drift 

<5 ppb / 24 hours

Response time

<60s

Linearity

<1%FS

Repeatability

<1%

Sampling flow range

900cm3/min±10%

Ambient temperature

5 ℃ ~ 40 ℃

Dimension

178 mm×430 mm×635 mm

Weight

<10kg

Power supply

200V~240V, 50/60Hz, 400W

Analog output

10V, 5V, 1V, 0.1V (optional)

Communication specification

RS232/RS485

Feature

1. Measurement range 0 ~ 500ppb to 0 ~ 10ppm; User selectable

2. Dark current compensation at fixed time, high stability

3. Multi-functional operation by microprocessor

4. Adaptive signal filtering technology to optimize the response time

5. High sensitivity, high repeatability, simple operation

6. Instrument parameters can be automatically monitored

7. Large capacity memory; automatic storage of historical data

8. Modular design, easy to replace and maintain